TM 10--3510--221--24
0159 00--2
STILL TEMPERATURE SENSOR REPLACEMENT --
Continued
0159 00
INSTALLATION
1. Connect electrical connector (1) to temperature sensor (2).
2. Access examine inputs function using maintenance menu (WP 0010 00), then observe and record temperature
reading for TE100.
3. Access temperature sensor calibration function using maintenance menu (WP 0010 00), then set output of
temperature sensor TE200 to same temperature reading as recorded for TE100.
4. Slide new nut (3), upper ferrule (7) and lower ferrule (5) onto new temperature sensor (2).
5. Insert temperature sensor (2) into fitting (4) at previously recorded dimension.
6. Tighten nut (3) onto fitting (4) to swage lower ferrule (5) on temperature sensor (2).
7. Loosen nut (3) and remove temperature sensor (2) from fitting (4).
8. Apply pipe thread sealant to threads of fitting (4).
9. Thread fitting (4) into still (6).
10. Insert temperature sensor (2) into fitting (4) then tighten nut (3).
11. Perform a HEAT TANKS operation (TM 10--3510--221--10). While water is boiling verify no steam leaks exist.
END OF WORK PACKAGE